BLINK FR is a new generation of fast response laser sensors (patented), based on a technology that enables to reach high speeds unreachable with standard thermopile detectors.
FAST RESPONSE LASER SENSORS
DESIGNED FOR OEM AND LABORATORY APPLICATIONS
Dozens of times faster
than standard Thermopile sensors
Power density capability:
1.5KW/cm2
Active area:
16x16mm
Wavelenght range:
Measures from 250nm to 1100nm and 10.6 microns
BLINK FR
A SUPERIOR TECHNOLOGY
BLINK FR sensors are based on a proprietary technology which enables natural response times of 90 ms (typ.) while still keeping broadband spectral range, power density capability and direct high power operation up to 60W of comparable thermopile detectors.
The response time is a key parameter for laser detectors where a faster response is required because several phenomena like pointing stability, fast drifts, instabilities can only be monitored if the detector responds very quickly. However, when the detector is fast it cannot withstand high powers (as in the case of all photodetectors) or, vice-versa, when the detector can withstand the power, it is slow (as it happens with thermopile sensors or Peltier devices).
Photodetectors can respond to pulses < 1ns. They can only withstand few milliwatts of direct laser power and their advantage of a fast response is, in most of applications, spoiled by standard readout electronics which curb the response time to 0.1-0.2 s.
Thermopile detectors are inherently slow: their natural response times range from > 1s to several tens of seconds, depending on the maximum power rating. Those natural response times can be reduced by electronics circuits and speed-up algorithms, but cannot be lower than 800ms-1 s, and only in the case of very low power heads.
BLINK FR sensors, differently from thermopile sensors, offer an unreached natural (without any acceleration by additional electronics) response time down to 90 ms (typ.). This is a game changing feature for laser power measurement because it paves the way to a number of applications that were not possible before.
Fast response laser sensors
Ordering Code | Power range | Energy range | Max Energy Density | Useful Aperture | Spectral Range | Absorber | Cooling | |
---|---|---|---|---|---|---|---|---|
BL-A-5W-16-K | 25 mW - 5W | 20 mJ – 5 J | 1 J/cm2 | 16x 16 mm | 0.2 – 25 µm | K | Convection | Ask for quote |
BL-A-8W-16-K | 25 mW - 8W | 20 mJ – 5 J | 1 J/cm2 | 16x 16 mm | 0.2 – 25 µm | K | Convection | Ask for quote |
BL-A-25W-16-K | 25 mW - 25W | 20 mJ – 5 J | 1 J/cm2 | 16x 16 mm | 0.2 – 25 µm | K | Forced Air | Ask for quote |
BL-W-50W-16-K | 30 mW - 50 W | 25 mJ – 5 J | 1 J/cm2 | 16 x 16 mm | 0.2 – 25 µm | K | Water | Ask for quote |
BL-A-5W-12-A | 50 mW - 5 W | 50 mJ – 10 J | 10 J/cm2 | 12 x 12 mm | 0.5 – 1.1 µm | A | Convection | Ask for quote |
BL-A-10W-12-A | 50 mW - 10 W | 50 mJ – 10 J | 10 J/cm2 | 12 x 12 mm | 0.5 – 1.1 µm | A | Convection | Ask for quote |
BL-A-50W-12-A | 100 mW - 50 W | 50 mJ – 10 J | 10 J/cm2 | 12 x 12 mm | 0.5 – 1.1 µm | A | Forced Air | Ask for quote |
BL-W-100W-12-A | 100 mW - 100 W | 50 mJ – 10 J | 10 J/cm2 | 12 x 12 mm | 0.5 – 1.1 µm | A | Water | Ask for quote |