Power Mode
Max. Average Power20 W
Max. Average Power (1)30 W
Min. Power40 mW
Power Resolution1 mW
Noise Equivalent Power (NEP)1 mW
Response Time (0-90%)3 sec
Power Calibration Uncertainty± 3%
Power Linearity (2)± 1%
Single Shot Energy Mode
Max. Energy (with 100 ms pulse)20 J
Min. Energy100 mJ
Energy Resolution1 mJ
Energy Calibration Uncertainty± 5%
Absorber Specs
Aperture40 mm
TypeBB
Absorber Spectral Range0.4 - 5.2 µm
Calibration Spectral Range0.4 - 5.2 µm
Max Power Density (3)35 W/cm²
Max Energy Density (3)Single Pulse: (a)
10ms pulse width: 13 J/cm²
<10µs pulse width: 10 J/cm²
General Characteristics
CoolingConvection
Weight0.9 kg
Dimension100 x 100 x 55 mm
Cable length - connector1.5 m - DB15
2.5 m – USB (U option)
1.5 m - RS232 (R option)
Stand and PostLight Duty Stand
Notes
(1). 2 minutes max
(2). Detector centrally irradiated @50% of useful surface.
(3). Damage thresholds also depend on power level. Please see damage graphs for more details.
(a). For repeated pulses, please see volume absorber damage graphs
 
Configurer les options

Étalonnage inclus (Cal-V)

Select Calibration Range
Check if compatible with
Calibration Spectral Range

Plage supplémentaire

Catégorie : Ordering code: 20-BB-D40
Lire termes et conditions