Power Mode | |
Max. Average Power | 5 W |
Max. Intermittent Power (1) | 50 W |
Min. Power | 50 mW |
Noise Equivalent Power (NEP) | 4 mW |
Natural Response Time (0-90%) | Typ. 90 ms (min. 50 ms – max. 120 ms) |
Power Calibration Uncertainty | ± 3% |
Power Linearity (2) | ± 3% |
Spatial uniformity (3) | ± 3% |
Energy Mode | |
Max. Energy | 10 J |
Min. Energy | 50 mJ |
Max repetition rate | 5 Hz |
Energy Calibration Uncertainty | ± 5% |
Absorber Specs | |
Aperture | 12 x 12 mm |
Type | A |
Absorber Spectral Range | 0.5 - 11 µm |
Calibration Spectral Range | 0.532 µm; 1.07 µm |
Max Power Density (4) | 10 KW /cm² (a) |
Max energy density (J/cm² ) | 10 ns pulse width: 10 |
General Characteristics | |
Cooling | Convection (b) |
Maximum sensor operating temperature | 60 °C |
Weight | 140 g |
Dimensions | 60 x 60 x 16.1 mm |
Cable length - connector | 1.5 m - bare wires 2.5 m – USB (U option) 1.5 m - RS232 (R option) |
Notes | |
(1) max 2 sec exposure, max 20% duty cycle. (2) Detector centrally irradiated @50% of useful surface. (3) 3mm beam diameter, scanning 80% of active area (4) Damage thresholds depend on power level. | (a) Measured at 1064 nm, 5 W. (b) Recommended ambient temperature: 10 -30 °C |
Configure options
Read the terms and Conditions