Power Mode | |
Max. Average Power | 200 W |
Min. Power | 0.5 W |
Power Resolution | 10 mW |
Noise Equivalent Power (NEP) | 10 mW |
Response Time (0-90%) | 5 sec |
Power Calibration Uncertainty | ± 3% |
Power Linearity (2) | ± 1% |
Absorber Specs | |
Aperture | 20 mm |
Type | HPB |
Absorber Spectral Range | 0.19 - 11 µm |
Calibration Spectral Range | 0.19 - 2.1 µm, 2.94µm, 9 - 11 µm |
Max Power Density (2) | 6 kW/cm² @200 W |
Max Energy Density (2) | 5ms pulse width: 36 J/cm² 10µs pulse width: 1.2 J/cm² 10ns pulse width: 0.3 J/cm² |
General Characteristics | |
Cooling | Conduction (a) |
Weight | 0.2 kg |
Dimension | Ø 56 x 21 mm |
Cable length - connector | 1.5 m – DB15) 2.5 m – USB (U option) 1.5 m - RS232 (R option) |
Notes | |
(1). Detector centrally irradiated @50% of useful surface. (2). Damage thresholds also depend on power level. Please see damage graphs for more details. | (a). Conduction, through heat sink |
Configure options
Read the terms and Conditions