Power Mode | |
Max. Average Power | 30 W |
Max. Average Power (1) | 45 W |
Min. Power | 20 mW |
Power Resolution | 1 mW |
Noise Equivalent Power (NEP) | 1 mW |
Response Time (0-90%) | 15 sec |
Power Calibration Uncertainty | ± 3% |
Power Linearity (2) | ± 1% |
Single Shot Energy Mode | |
Max. Energy (with 100 ms pulse) | 45 J |
Min. Energy | 100 mJ |
Energy Resolution | 1 mJ |
Energy Calibration Uncertainty | ± 5% |
Absorber Specs | |
Aperture | 12 mm |
Type | SHC + L |
Absorber Spectral Range | 0.2 - 1.1 µm |
Calibration Spectral Range | 0.2 - 1.1 µm |
Max Power Density (3) | 100 kW/cm² |
Max Energy Density (3) | 5ms pulse width: 320 J/cm² 10µs pulse width: 12 J/cm² 10ns pulse width: 3 J/cm² |
General Characteristics | |
Cooling | Convection |
Weight | 0.6 kg |
Dimension | 75 x 75 x 73 mm |
Cable lenght - connector | 1.5 m - DB15 2.5 m – USB (U option) 1.5 m - RS232 (R option) |
Stand and Post | Light Duty Stand |
Notes | |
(1). 2 minutes max (2). Detector centrally irradiated @50% of useful surface. (3). Damage thresholds also depend on power level. Please see damage graphs for more details |
Seçenekleri yapılandır
Şartlar ve Koşulları okuyun